FDA Enforcement Class II Terminated

7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY; 9 Hole VL Gridlock Fibula Plate, Part # 300-60-002 NON-STERILE SINGLE USE ONLY; 11 Hole VL Gridlock Fibula Plate, Part # 300-60-003 NON-STERILE SINGLE USE ONLY. For use in trauma and reconstructive procedures in small bones.

Recall: Z-2490-2015 · Reported September 9, 2015

Enforcement

Recall Number
Z-2490-2015
Event ID
71967
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Trilliant Surgical Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 9, 2015
Initiation Date
August 6, 2015
Classification Date
August 31, 2015
Termination Date
January 26, 2017
Address
6721 Portwest Dr Ste 160, N/A, Houston, TX, 77024-8019, United States

Description

7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY; 9 Hole VL Gridlock Fibula Plate, Part # 300-60-002 NON-STERILE SINGLE USE ONLY; 11 Hole VL Gridlock Fibula Plate, Part # 300-60-003 NON-STERILE SINGLE USE ONLY. For use in trauma and reconstructive procedures in small bones.

Reason

The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.

Code Info

Lots: TSL002595, TSL002596, TSL002597

Distribution

US Distribution to the states of : FL, OH, NJ, MN, WI, NM and CA.

Quantity

28 units