FDA Enforcement Class II Terminated

Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.

Recall: Z-2487-2015 · Reported September 9, 2015

Enforcement

Recall Number
Z-2487-2015
Event ID
71988
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Pointe Scientific, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 9, 2015
Initiation Date
May 19, 2015
Classification Date
August 31, 2015
Termination Date
May 23, 2016
Address
5449 Research Dr, N/A, Canton, MI, 48188-2261, United States

Description

Pointe Scientific G6PD Controls Kit configuration: 6 x 0.5 ml (2 vials each of Level 1, Level 2 and Level 3). Also sold as individual vials Bottle: Clear glass vials Cap: Rubber seal with colored aluminum seals Glucose-6-phosphate dehydrogenase (G6PD) controls can be used to test for the quantitative and qualitative determination of G6PD in blood.

Reason

A reduction in the reconstituted stability has been identified. Clinicians may notice a drop in recovery of the control after reconstitution, which could result in an out of specification result prior to the current reconstitution claim of 7 days. For this reason we suggest discontinuing use of the control set.

Code Info

Lot: 505801 Expiration date: 2015-11 catalog numbers: G7583-CTL 7-G7583-CTL-L1 7-G7583-CTL-L2 7-G7583-CTL-L3

Distribution

Worldwide distribution: US (nationwide) including states of: PA, CA, KY, FL, NJ, DE, MI, MN,and IL; and countries of: United Arab Emirates, Austria, Italy, Israel, Greece, and Australia.

Quantity

2772 total vials (374 kits of 6 vials, 528 invidually sold vials)