RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.
Enforcement
- Recall Number
- Z-2486-2025
- Event ID
- 97311
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Reflexion Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 10, 2025
- Initiation Date
- July 22, 2025
- Classification Date
- September 2, 2025
- Address
- 25881 Industrial Blvd, N/A, Hayward, CA, 94545-2991, United States
Description
RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
Model Number: RXM1000 UDI Codes that includes Serial Numbers: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012
U.S. Nationwide distribution in the states of CA, CT, LA, NJ, OR, PA, and TX.
11