Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottle, 2.5mL. Indicated for use with fresh fingertip capillary whole blood samples.
Enforcement
- Recall Number
- Z-2482-2015
- Event ID
- 71611
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Bayer Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 9, 2015
- Initiation Date
- June 22, 2015
- Classification Date
- August 28, 2015
- Termination Date
- May 23, 2016
- Address
- 430 S Beiger St, N/A, Mishawaka, IN, 46544-3207, United States
Description
Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottle, 2.5mL. Indicated for use with fresh fingertip capillary whole blood samples.
If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.
SKU# 7314 (Level 2) and SKU# 7315 (Level 1) All Lots
Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Slovenia, Spain, Sweden, Switzerland, UK, Albania, Azerbaijan, Bulgaria, Czech Republic, Estonia, Hungary, Korea, Latvia, Lithuania, Philippines, Romania, Serbia, Slovakia, and Thailand.
5 million bottles of control solution