MEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
Enforcement
- Recall Number
- Z-2480-2026
- Event ID
- 98994
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 24, 2026
- Initiation Date
- May 4, 2026
- Classification Date
- June 17, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
MEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
1. DYNC1816, UDI/DI 10080196990627 (each), 40080196990628 (case); 2. DYND10350, UDI/DI 10080196028665 (each), 40080196028666 (case). ALL LOTS
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
8327756 units