FDA Enforcement
Class II
Ongoing
Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.
Recall: Z-2478-2025
·
Reported September 17, 2025
Enforcement
- Recall Number
- Z-2478-2025
- Event ID
- 97520
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- WASATCH PHOTONICS
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- N/A
- Report Date
- September 17, 2025
- Initiation Date
- August 4, 2025
- Classification Date
- September 11, 2025
- Address
- 808 Aviation Pkwy Ste 1400, N/A, Morrisville, NC, 27560-6600, United States
Description
Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.
Reason
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
Code Info
all codes included in the model numbers.
Distribution
U.S.
Quantity
192