Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500
Recall: Z-2477-2026
·
Reported June 24, 2026
Enforcement
- Recall Number
- Z-2477-2026
- Event ID
- 98986
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- GE Medical Systems, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 24, 2026
- Initiation Date
- May 1, 2026
- Classification Date
- June 17, 2026
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615, United States
Description
Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500
Reason
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Code Info
Serial Numbers: 5452896 801057486 800977619 801093759 5584569 5587625 5367527 5603110 5371240 5497382
Distribution
Worldwide - US Nationwide distribution.
Quantity
265 units