FDA Enforcement
Class II
Ongoing
epoc BGEM BUN Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Recall: Z-2476-2023
·
Reported September 6, 2023
Enforcement
- Recall Number
- Z-2476-2023
- Event ID
- 92537
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 6, 2023
- Initiation Date
- June 1, 2023
- Classification Date
- August 28, 2023
- Address
- 2 Edgewater Dr, N/A, Norwood, MA, 02062-4637, United States
Description
epoc BGEM BUN Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Reason
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Code Info
UDI-DI: 00809708121860 Siemens Material Number: 10736515 Lots: 04-23003-60 (expiry 20-Jun-2023), 11-22337-60 (expiry 20-May-2023), 04-23034-40 (expiry 21-Jul-2023), 04-23059-50 (expiry 15-Aug-2023), 04-23059-60 (expiry 15-Aug-2023), and 04-23069-40 (expiry 25-Aug-2023)
Distribution
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam
Quantity
3201 ea