FDA Enforcement Class I Terminated

GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA: The Core OneTouch Cables (Reusable Device) is ancillary equipment used with the Core 10 and Core 15 monitors. The monitor and workstation are intended to work with video endoscopes, in conjunction with ancillary equipment, for endoscopic procedures.

Recall: Z-2476-2020 · Reported July 15, 2020

Enforcement

Recall Number
Z-2476-2020
Event ID
85705
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Verathon, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 15, 2020
Initiation Date
June 5, 2020
Classification Date
July 9, 2020
Termination Date
April 22, 2024
Address
20001 N Creek Pkwy, Bothell, WA, 98011-8218, United States

Description

GlideScope Core OneTouch Smart Cable, REF: 0800 0601, an accessory to the GlideScope Video Laryngoscope System Per PLRA: The Core OneTouch Cables (Reusable Device) is ancillary equipment used with the Core 10 and Core 15 monitors. The monitor and workstation are intended to work with video endoscopes, in conjunction with ancillary equipment, for endoscopic procedures.

Reason

When video laryngoscopy system users apply, twisting motions while connecting the blade in the HDMI port, or excessive torsion to the connection port, this force is transferred to the HDMI PCB, which may result in video synchronization interruption, loss of live image/blank screen, no cameras connected monitor warning, may present risk to patients, in which time to intubation is more critical.

Code Info

All serial numbers

Distribution

Worldwide Distribution: US (nationwide) and countries of: Australia and United Kingdom.

Quantity

437