FDA Enforcement Class II Terminated

Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - Product Usage: The Alere Cholestech LDX System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX¿ Analyzer.

Recall: Z-2475-2019 · Reported September 11, 2019

Enforcement

Recall Number
Z-2475-2019
Event ID
83410
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Alere San Diego, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
September 11, 2019
Initiation Date
February 15, 2019
Classification Date
September 5, 2019
Termination Date
December 12, 2023
Address
9975 Summers Ridge Rd, N/A, San Diego, CA, 92121-2997, United States

Description

Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - Product Usage: The Alere Cholestech LDX System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX¿ Analyzer.

Reason

The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.

Code Info

Lot/Unit Codes: Lot codes of Meter Kits: 418197, 418198, 418199, 418252, & 418253 Lot codes of Power Supplies: 724599, 724604, & 725914 Serial Numbers: AA184455, AA184344, AA184328, AA184388, AA184304, AA184349, AA184396, AA184339, AA184421, AA184280, AA184435, AA184427, AA184395, AA183401, AA184394, AA184710, AA184345, AA184039, AA184463, AA184609, AA184383, AA184415, AA184094, AA184070, AA184412, AA183861, AA184278, AA184283, AA184711, AA184392, AA184343, AA184368, AA184708, AA184350, AA184674, AA184378, AA184433, AA184366, AA184365, AA184354, AA184369, AA184067, AA184353, AA184664, AA184279, AA184355, AA184270, AA184285, AA184017, AA184272, AA184340, AA184346, AA184347, AA184336, AA184342, AA184269, AA184376, AA184949, AA184386, AA184351, AA184341, AA184413, AA184298, AA183707, AA184595, AA184364, AA184287, AA184618, AA184971, AA184334, AA184845, AA184267, AA184296, AA184333, AA184608, AA184651, AA184319, AA184348, AA184317, AA184699, AA184430, AA184432, AA184363, AA184030, AA184338, AA184371, AA184375, AA184393, AA184306, AA184447, AA183418, AA184323, AA184401, AA184438, AA184594, AA184281, AA184461, AA184312, AA184321, AA184288, AA184357, AA184380, AA184617, AA184352, AA184273, AA184303, AA184424, AA184290, AA184300, AA184916, AA184965, AA184963, AA184980, AA184964, AA184955, AA185069, AA185009, AA184904, AA184993, AA184973, AA184991, AA184995, AA184900, AA185046, AA184951, AA185927, AA184468, AA185050, AA184967, AA184853, AA184877, AA184830, AA184961, AA184794, AA184930, AA184459, AA184887, AA184869, AA185020, AA184785, AA184985, AA184901, AA184979, AA184990, AA184865, AA184988, AA184872, AA184952, AA184946, AA184924, AA182949, AA184918, AA184928, AA184886, AA184957, AA184974, AA184862, AA185022, AA184657, AA184880, AA185021, AA184922, AA184875, AA185013, AA184943, AA184873, AA184989, AA184742, AA184876, AA184301, AA184367, AA184370, AA184359, AA184434, AA184615, AA185004, AA184678, AA184329, AA184377, AA184326, AA184284, AA184384, AA184795, AA184297, AA183549, AA184318

Distribution

US Nationwide distribution in the states of AK, A, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY< OH, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI. No OUS consignees

Quantity

186 analyzer kits