FDA Enforcement Class II Terminated

The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. "Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). "Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures."

Recall: Z-2474-2021 · Reported September 22, 2021

Enforcement

Recall Number
Z-2474-2021
Event ID
88546
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips North America Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 22, 2021
Initiation Date
August 13, 2021
Classification Date
September 14, 2021
Termination Date
March 11, 2024
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States

Description

The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. "Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). "Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures."

Reason

Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.

Code Info

Serial numbers: 48, 66, 67, 160

Distribution

US Nationwide distribution in the states of ID, MA, MI, TX, VA.

Quantity

4 units