FDA Enforcement
Class II
Ongoing
uCT 530 Computed Tomography X-Ray System
Recall: Z-2472-2020
·
Reported July 8, 2020
Enforcement
- Recall Number
- Z-2472-2020
- Event ID
- 85809
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 8, 2020
- Initiation Date
- May 8, 2020
- Classification Date
- June 30, 2020
- Address
- No. 2258 Chengbei Rd, Jiading Ind., N/A, Shanghai, N/A, N/A, China
Description
uCT 530 Computed Tomography X-Ray System
Reason
Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.
Code Info
Units with software version no. R001.11.0.729741-Re-1100 including the following: Serial number/GTIN: 300162 (GTIN: 6971576831012); 300163 (GTIN: 6971576831012); 300174 (GTIN: 6971576831012); and 300176 (GTIN: 6971576831012).
Distribution
Distributed to Texas.
Quantity
4 units