uCT 530 Computed Tomography X-Ray System

Recall: Z-2472-2020 · Reported July 8, 2020

Enforcement

Recall Number
Z-2472-2020
Event ID
85809
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Shanghai United Imaging Healthcare Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 8, 2020
Initiation Date
May 8, 2020
Classification Date
June 30, 2020
Address
No. 2258 Chengbei Rd, Jiading Ind., N/A, Shanghai, N/A, N/A, China

Description

uCT 530 Computed Tomography X-Ray System

Reason

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Code Info

Units with software version no. R001.11.0.729741-Re-1100 including the following: Serial number/GTIN: 300162 (GTIN: 6971576831012); 300163 (GTIN: 6971576831012); 300174 (GTIN: 6971576831012); and 300176 (GTIN: 6971576831012).

Distribution

Distributed to Texas.

Quantity

4 units