FDA Enforcement Class II Ongoing

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G;

Recall: Z-2465-2025 · Reported September 3, 2025

Enforcement

Recall Number
Z-2465-2025
Event ID
97358
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 3, 2025
Initiation Date
July 14, 2025
Classification Date
August 28, 2025
Address
85 Shaffer Park Dr, Tiffin, OH, 44883-9290, United States

Description

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G;

Reason

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Code Info

1) Model/Item Number CILC69L: UDI/DI 00191072235503, Kit Lot Numbers: 42-7937111, 42-8049311; 2) Model/Item Number HNLC80J: UDI/DI 00191072193650, Kit Lot Numbers: 42-8044511, 42-8118911, 42-7977011, 42-7803211; 3) Model/Item Number JKLC55H: UDI/DI 00191072235138, Kit Lot Numbers: 42-7814311; 4) Model/Item Number LLLA21K: UDI/DI 00191072233486, Kit Lot Numbers: 70-052582, 70-053048, 70-051564, 70-051373, 70-051163, 70-053446, 70-053776; 5) Model/Item Number MPGL15G: UDI/DI 00191072223197, Kit Lot Numbers: 5436916, 5436615

Distribution

US distribution to states of: IL and LA

Quantity

1233 units