FDA Enforcement Class II Terminated

HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: 2015-08-13 Container Type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2-8C on receipt. Light Sensitive. In vitro diagnostic www.HardyDiagnostics.com

Recall: Z-2463-2015 · Reported September 2, 2015

Enforcement

Recall Number
Z-2463-2015
Event ID
71804
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hardy Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
September 2, 2015
Initiation Date
July 9, 2015
Classification Date
August 26, 2015
Termination Date
November 10, 2015
Address
1430 W Mccoy Ln, N/A, Santa Maria, CA, 93455-1005, United States

Description

HardyCHROM MRSA Cat no: G307 Lot no: H15155 Expires: 2015-08-13 Container Type: 15x100mm monoplate Packaged: 10 plates/sleeve Storage: 2-8C on receipt. Light Sensitive. In vitro diagnostic www.HardyDiagnostics.com

Reason

Hardy Diagnostics is recalling HardyCHROm MRSA due to a report of the potential for breakthrough of the methicillin-sensitive Staphylococcus aureus (MSSA) QC strain, ATCC 29213 on the identified lot.

Code Info

Cat no. G307 Lot no. H15155

Distribution

Distributed in the states of VA, PA, NY, NJ, IL, OH, and MD.

Quantity

2,820 (282 pk/10)