Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.
Enforcement
- Recall Number
- Z-2461-2015
- Event ID
- 71556
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ferno-Washington Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 2, 2015
- Initiation Date
- May 1, 2015
- Classification Date
- August 25, 2015
- Termination Date
- November 6, 2015
- Address
- 70 Weil Way, N/A, Wilmington, OH, 45177-9371, United States
Description
Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.
The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.
Serial numbers: 15F001703, 15F001704, 15F001705, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001714, 15F001715, 15F001716, 15F001717, 15F001719, 15F001720, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001729 & 15F001730.
Distributed in the states of AZ, CO, IA, MD, OH, PA & WA, and the countries of Australia, Canada & United Kingdom.
26 units