FDA Enforcement
Class II
Terminated
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.
Recall: Z-2457-2015
·
Reported September 2, 2015
Enforcement
- Recall Number
- Z-2457-2015
- Event ID
- 71414
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- RAYSEARCH LABORATORIES AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- September 2, 2015
- Initiation Date
- May 27, 2015
- Classification Date
- August 25, 2015
- Termination Date
- September 22, 2017
- Address
- Sveavaegen 9, N/A, Stockholm, N/A, N/A, Sweden
Description
RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning. Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.
Reason
Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.
Code Info
software build numbers are 4.7.1.10
Distribution
Distributed in CO, IL, MI, NC, OH,TN, and WA.
Quantity
126 programs