FDA Enforcement Class I Ongoing

IDM-MICRO-R. Electrical wheelchair component.

Recall: Z-2456-2025 · Reported September 10, 2025

Enforcement

Recall Number
Z-2456-2025
Event ID
97354
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
mo-Vis BVBA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 10, 2025
Initiation Date
July 10, 2025
Classification Date
September 2, 2025
Address
Biebuyckstraat 15D, N/A, Nevele, N/A, N/A, Belgium

Description

IDM-MICRO-R. Electrical wheelchair component.

Reason

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Code Info

Model No. PRSPS0014; UDI-DI: 05407008320836; Serial No. 1000 to 1690.

Distribution

US Nationwide distribution in the state of TX.

Quantity

15 units