FDA Enforcement Class II Ongoing

BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853

Recall: Z-2456-2024 · Reported August 7, 2024

Enforcement

Recall Number
Z-2456-2024
Event ID
94886
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
BioFire Diagnostics, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 7, 2024
Initiation Date
June 19, 2024
Classification Date
July 31, 2024
Address
515 S Colorow Dr, N/A, Salt Lake City, UT, 84108-1248, United States

Description

BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853

Reason

If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.

Code Info

BCID2 Panel REF: RFIT-ASY-0147, UDI-DI: 00815381020338, All unexpired lots. BACT/ALERT REF/UDI-DI: Lot/Expiration: 410851/03573026596057: 0004101638/3/20/2024, 0004101930/7/28/2024, 0004102026/9/4/2024, 0004102408/2/16/2025, 0004102996/9/24/2025, 0004102946/8/29/2025, 0004102956/9/17/2025, 0004102998/9/24/2025. 410853/03573026596095: 0004101718/4/24/2024, 0004101790/6/1/2024, 0004101958/8/3/2024, 0004102452/3/3/2025, 0004102964/9/19/2025

Distribution

Worldwide - US Nationwide distribution including in the states of FL, IL, KY, CO, KS, NC, AK, NY, ND, CA, TX, MD, GA, OH, WI, SD, LA, MT, VA, IA, MO, MA, WV, NE, OK, AZ, OR, SC, MI, IN, ID, TN, WA, NM, NJ, AR, AL, MN, NH, NV, PA, WY, MS, CT and the countries of AE, AM, AR, AT, BA, BG, BH, BN, BR, CA, CH, CI, CK, CN, CO, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GA, GB, GE, GR, GT, HR, HU, IL, IN, IQ, IS, IT, JO, JP, KK, KR, KW, LB, LT, LV, MK, MM, MX, MY, NG, NO, NW, OM, PA, PH, PK, PL, PT, QA, RO, RS, SA, SI, TN, TR, TW, UG, UY, VN, WF.

Quantity

99,822