FDA Enforcement
Class I
Ongoing
Multi Joystick R-net. Electrical wheelchair component.
Recall: Z-2455-2025
·
Reported September 10, 2025
Enforcement
- Recall Number
- Z-2455-2025
- Event ID
- 97354
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- mo-Vis BVBA
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- September 10, 2025
- Initiation Date
- July 10, 2025
- Classification Date
- September 2, 2025
- Address
- Biebuyckstraat 15D, N/A, Nevele, N/A, N/A, Belgium
Description
Multi Joystick R-net. Electrical wheelchair component.
Reason
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Code Info
Model No. P002-62; UDI-DI: 05407008320041; Serial No. 1000 to 1690.
Distribution
US Nationwide distribution in the state of TX.
Quantity
3 units