FDA Enforcement Class II Terminated

Sterile Disposable Connectors ASY S/1/4 TO 3/16 RED 24 REF/ GTIN for Insert Label:050514000/00803622123269

Recall: Z-2450-2020 · Reported July 8, 2020

Enforcement

Recall Number
Z-2450-2020
Event ID
85623
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
LivaNova USA Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 8, 2020
Initiation Date
April 24, 2020
Classification Date
June 26, 2020
Termination Date
February 17, 2022
Address
14401 W 65th Way, N/A, Arvada, CO, 80004-3503, United States

Description

Sterile Disposable Connectors ASY S/1/4 TO 3/16 RED 24 REF/ GTIN for Insert Label:050514000/00803622123269

Reason

The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.

Code Info

LOT 2004300006

Distribution

US - AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV OUS- Spain (Hospital de Barsuto, Bilbao, Spain)

Quantity

96 units