FDA Enforcement Class I Ongoing

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used of each are the same for each user. The watchOS app cannot be installed independently from the iOS phone app Component: Dexcom G7 CGM iOS App

Recall: Z-2448-2025 · Reported September 10, 2025

Enforcement

Recall Number
Z-2448-2025
Event ID
97262
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Dexcom, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 10, 2025
Initiation Date
July 24, 2025
Classification Date
September 4, 2025
Address
6340 Sequence Dr, N/A, San Diego, CA, 92121-4356, United States

Description

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used of each are the same for each user. The watchOS app cannot be installed independently from the iOS phone app Component: Dexcom G7 CGM iOS App

Reason

The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Code Info

Lot Code: UDI - SW1355- 00386270005168

Distribution

SW12299 G7 Android CGM App Worldwide Distribution - US Nationwide and the countries of Andorra, United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW12300 G7 iOS CGM App Worldwide Distribution - US Nationwide and the countries of United Arab Emirates, Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, South Korea, Kuwait, Lebanon, Liechtenstein, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa SW14245 Dexcom ONE+ Android Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Israel, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden SW14244 D1G7 iOS Mobile App OUS worldwide distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Estonia, Spain, France, United Kingdom, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, Sweden

Quantity

907842