FDA Enforcement Class II Ongoing

DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE GLOSS 680 ROLLATOR 1655081 - DOLOMITE GLOSS 600 ROLLATOR 1655082 - DOLOMITE GLOSS 520 ROLLATOR 1655083 - DOLOMITE GLOSS 450 ROLLATOR

Recall: Z-2445-2023 · Reported August 30, 2023

Enforcement

Recall Number
Z-2445-2023
Event ID
92645
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Invacare Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 30, 2023
Initiation Date
June 15, 2023
Classification Date
August 23, 2023
Address
1200 Taylor St, N/A, Elyria, OH, 44035-6248, United States

Description

DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE GLOSS 680 ROLLATOR 1655081 - DOLOMITE GLOSS 600 ROLLATOR 1655082 - DOLOMITE GLOSS 520 ROLLATOR 1655083 - DOLOMITE GLOSS 450 ROLLATOR

Reason

Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user

Code Info

Serial number in the range: 20BHL0002 to 21IHL1910

Distribution

US Nationwide distribution.

Quantity

586 devices