FDA Enforcement Class II Terminated

cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001

Recall: Z-2444-2021 · Reported September 22, 2021

Enforcement

Recall Number
Z-2444-2021
Event ID
88398
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Diagnostics Operations, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 22, 2021
Initiation Date
June 24, 2021
Classification Date
September 10, 2021
Termination Date
July 21, 2023
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001

Reason

Under specific circumstances created by the user, the cobas e flow test results could be replaced by an automatic result sent to the lab LIS. This automatic result could be released and interpreted by your Laboratory Information System (LIS) as a false positive or negative result. It should be noted that the "correct" result would still be available on the instrument.

Code Info

UDI: 04015630936007; All systems running software version 3.01.03 - 3.02.08

Distribution

US Nationwide distribution

Quantity