FDA Enforcement
Class II
Terminated
cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001
Recall: Z-2444-2021
·
Reported September 22, 2021
Enforcement
- Recall Number
- Z-2444-2021
- Event ID
- 88398
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 22, 2021
- Initiation Date
- June 24, 2021
- Classification Date
- September 10, 2021
- Termination Date
- July 21, 2023
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001
Reason
Under specific circumstances created by the user, the cobas e flow test results could be replaced by an automatic result sent to the lab LIS. This automatic result could be released and interpreted by your Laboratory Information System (LIS) as a false positive or negative result. It should be noted that the "correct" result would still be available on the instrument.
Code Info
UDI: 04015630936007; All systems running software version 3.01.03 - 3.02.08
Distribution
US Nationwide distribution
Quantity
2058 licenses worldwide, 79 licenses in US