FDA Enforcement Class II Terminated

AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Recall: Z-2444-2019 · Reported September 11, 2019

Enforcement

Recall Number
Z-2444-2019
Event ID
83334
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Angiodynamics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
September 11, 2019
Initiation Date
June 27, 2019
Classification Date
August 30, 2019
Termination Date
July 29, 2020
Address
603 Queensbury Ave, N/A, Queensbury, NY, 12804-7619, United States

Description

AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Reason

The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.

Code Info

Lot 5456490; UPN H787114030120

Distribution

The products were distributed to the following US states: DC, IL, IN, MI, NJ, TN, TX, and WI.

Quantity

109