Terumo Advanced Perfusion System 1 ASSEMBLY FLOW POD Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Enforcement
- Recall Number
- Z-2443-2012
- Event ID
- 62313
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Terumo Cardiovascular Systems Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 3, 2012
- Initiation Date
- September 10, 2012
- Classification Date
- September 23, 2012
- Termination Date
- January 13, 2014
- Address
- 6200 Jackson Road, N/A, Ann Arbor, MI, 48103-9586, United States
Description
Terumo Advanced Perfusion System 1 ASSEMBLY FLOW POD Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Terumo Cardiovascular Systems (CVS) received two reports of the false back flow alarms for the Terumo¿ Advanced Perfusion System 1 when there was no tubing in the flow sensor. Terumo CVS' investigation found that the voltage level on a circuit board was improperly set in a population of flow modules for Terumo System 1, the sensor operates correctly if the tubing is installed and it is highly unli
Catalog number: 802018 and serial numbers: 00032, 00042, 00056, 00147, 00623, 00700-00764, 00766-00820, and 00823-00847.
Worldwide and US Nationwide Distribution
150 units