FDA Enforcement Class II Terminated

Terumo Advanced Perfusion System 1 ASSEMBLY FLOW POD Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Recall: Z-2443-2012 · Reported October 3, 2012

Enforcement

Recall Number
Z-2443-2012
Event ID
62313
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 3, 2012
Initiation Date
September 10, 2012
Classification Date
September 23, 2012
Termination Date
January 13, 2014
Address
6200 Jackson Road, N/A, Ann Arbor, MI, 48103-9586, United States

Description

Terumo Advanced Perfusion System 1 ASSEMBLY FLOW POD Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Reason

Terumo Cardiovascular Systems (CVS) received two reports of the false back flow alarms for the Terumo¿ Advanced Perfusion System 1 when there was no tubing in the flow sensor. Terumo CVS' investigation found that the voltage level on a circuit board was improperly set in a population of flow modules for Terumo System 1, the sensor operates correctly if the tubing is installed and it is highly unli

Code Info

Catalog number: 802018 and serial numbers: 00032, 00042, 00056, 00147, 00623, 00700-00764, 00766-00820, and 00823-00847.

Distribution

Worldwide and US Nationwide Distribution

Quantity

150 units