FDA Enforcement
Class II
Ongoing
Medline convenience kits: FEMORAL BLOCK TRAY DYNJRA1739B
Recall: Z-2442-2026
·
Reported June 24, 2026
Enforcement
- Recall Number
- Z-2442-2026
- Event ID
- 98951
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 24, 2026
- Initiation Date
- April 27, 2026
- Classification Date
- June 12, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline convenience kits: FEMORAL BLOCK TRAY DYNJRA1739B
Reason
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Code Info
DYNJRA1739B UDI-DI 10195327411237 Lot 25LBB731
Distribution
US Nationwide distribution. OUS distribution pending.
Quantity
49,654 kits total