Health Harmony Mobile application software Product Usage: Care Innovations Health Harmony Mobile is intended as a communication tool to display medical device data from third party devices for patients to view, and to collect assessment (question & answers) from patients in the home.Care Innovations Health Harmony Mobile is intended as a communication tool to display medical device data from third party devices for patients to view, and to collect assessment (question & answers) from patients in the home. Patients can review their stored vital sign measurements (captured outside of Health Harmony Mobile) and receive educational and motivational content from caregivers.
Enforcement
- Recall Number
- Z-2442-2018
- Event ID
- 80382
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Intel-GE Care Innovations LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 25, 2018
- Initiation Date
- December 6, 2017
- Classification Date
- July 16, 2018
- Termination Date
- October 23, 2020
- Address
- 3721 Douglas Blvd Ste 100, Roseville, CA, 95661-4243, United States
Description
Health Harmony Mobile application software Product Usage: Care Innovations Health Harmony Mobile is intended as a communication tool to display medical device data from third party devices for patients to view, and to collect assessment (question & answers) from patients in the home.Care Innovations Health Harmony Mobile is intended as a communication tool to display medical device data from third party devices for patients to view, and to collect assessment (question & answers) from patients in the home. Patients can review their stored vital sign measurements (captured outside of Health Harmony Mobile) and receive educational and motivational content from caregivers.
It was discovered that in certain situations, including partial sessions and when taking adhoc measurements, the patient data was not synchronizing in a timely manner with the backend database, resulting in the patient's clinician not getting patient data tor one or two days.
Version v 2.5.2.29
US Nationwide Distribution in the states of CO, ID, PA, and LA