FDA Enforcement
Class II
Terminated
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Recall: Z-2441-2018
·
Reported July 25, 2018
Enforcement
- Recall Number
- Z-2441-2018
- Event ID
- 80418
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zeiss, Carl Inc
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- July 25, 2018
- Initiation Date
- May 29, 2018
- Classification Date
- July 19, 2018
- Termination Date
- July 26, 2024
- Address
- 1 Zeiss Dr, Thornwood, NY, 10594-1939, United States
Description
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Reason
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Code Info
432332-9020-000
Distribution
US Nationwide Distribution
Quantity
8