Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993
Enforcement
- Recall Number
- Z-2439-2020
- Event ID
- 85505
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- LivaNova Deutschland GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 8, 2020
- Initiation Date
- August 26, 2019
- Classification Date
- June 26, 2020
- Termination Date
- April 22, 2021
- Address
- Lindberghstr. 25, N/A, Munich, N/A, N/A, Germany
Description
Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993
The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.
Serial Number: 60S14735 60S14877 60S14715 60S14716 60S14736 60S14755 60S14707 60S14708
US: FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA OUS: Canada - Uzbekistan - Turkey - Tunisia - Taiwan - South Africa - Serbia - Saudi Arabia - Russia - Poland - Mexico - Korea(Rep. of) - India - Hong Kong - Colombia - China - Brazil - Bahrain - Argentina - Algeria - Italy - Germany