FDA Enforcement Class II Ongoing

Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Product Description: Perfusor Space Syringe Pump, Non-Wireless, Infusion Pump, 1 each Component: N/A

Recall: Z-2437-2024 · Reported July 31, 2024

Enforcement

Recall Number
Z-2437-2024
Event ID
94890
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
B Braun Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 31, 2024
Initiation Date
June 14, 2024
Classification Date
July 25, 2024
Address
200 Boulder Dr, Breinigsville, PA, 18031-1532, United States

Description

Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Product Description: Perfusor Space Syringe Pump, Non-Wireless, Infusion Pump, 1 each Component: N/A

Reason

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Code Info

Lot Code: Model No 8713032U; UDI-DI 04046964797965; All Serial Numbers

Distribution

Domestic distribution nationwide. International distribution to Canada.

Quantity

4519