FDA Enforcement Class I Ongoing

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Recall: Z-2436-2023 · Reported September 6, 2023

Enforcement

Recall Number
Z-2436-2023
Event ID
92491
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Datascope Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 6, 2023
Initiation Date
July 31, 2023
Classification Date
August 30, 2023
Address
1300 Macarthur Blvd, N/A, Mahwah, NJ, 07430-2052, United States

Description

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Reason

Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kinks/abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The firm's HHE also identified hardware issues related to these alarms.

Code Info

All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449

Distribution

Worldwide - US Nationwide distribution.

Quantity

9175 units