Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.
Enforcement
- Recall Number
- Z-2435-2021
- Event ID
- 88411
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- NIHON KOHDEN ORANGEMED, INC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 15, 2021
- Initiation Date
- July 20, 2021
- Classification Date
- September 8, 2021
- Termination Date
- April 21, 2023
- Address
- 1800 E Wilshire Ave, N/A, Santa Ana, CA, 92705-4617, United States
Description
Nihon Kohden NKV-550 Ventilator, REF: NKV-550-U, CE 0344, UDI: (01) 00843685100012 The Nihon Kohden NKV-550 Series Ventilator System is intended to provide continuous ventilation for adult, pediatric and neonatal patients who require invasive or noninvasive respiratory support.
Potential defective gas inlet pressure regulator within ventilator which could result in gas releasing noise from the bottom of the ventilator, a High Air/O2 Inlet Pressure alarm, a High/Low O2 alarm, or could not pass device Check prior to use.
All serial numbers between NKV5502019xxxx - NKV5502112xxxx and the following serial numbers below: NKV55019270055, NKV55019370098R, NKV55019430125R, NKV55019470156R, NKV55019470160, NKV55020130200, NKV55020140213R, NKV55020140226R, NKV55020140229, NKV55020150241, NKV55020150252, NKV55020160268, NKV55020160276, NKV55020160277, NKV55020160292R, NKV55020160296, NKV55020160297, NKV55020160304, NKV55020160307, NKV55020160308, NKV55020160310, NKV55020160312, NKV55020170317, NKV55020170318, NKV55020170325, NKV55020170335, NKV55020170336, NKV55020170341, NKV55020170342, NKV55020170343, NKV55020170346, NKV55020170347, NKV55020170349, NKV55020170350, NKV55020170352, NKV55020170354, NKV55020170357, NKV55020170360, NKV55020180365, NKV55020180366, NKV55020180369, NKV55020180370, NKV55020180371, NKV55020180372, NKV55020180373, NKV55020180374, NKV55020180375, NKV55020180376, NKV55020180378, NKV55020180379, NKV55020180381, NKV55020180383, NKV55020180384, NKV55020180385, NKV55020180387, NKV55020180391, NKV55020180392, NKV55020180396, NKV55020180397, NKV55020180398, NKV55020180399, NKV55020180400, NKV55020180401, NKV55020180402, NKV55020180403, NKV55020180405, NKV55020180406, NKV55020180408, NKV55020180409, NKV55020180410, NKV55020180411, NKV55020180412, NKV55020180413
Worldwide Distribution: US (nationwide) to states of: AZ, CA, NY, MA, VT, LA, DC, PA, FL, UT and OUS (countries) of: Brazil, Japan, Ireland, Poland, Scotland, Germany, Kazakhstan, Spain, Slovenia, Germany/Bulgaria, Russia, India, Palestine, Libya, Jordan Kuwait, Lebanon, Syria, Turkey, South Africa, United Arab Emirates, Mexico, Colombia, Peru, Costa Rica, Barbados, Ecuador, Bahamas, Bolivia, St Kitts & Nevis, St. Martin, Chile, Thailand, Laos, Philippines, Malaysia, Nepal and United Kingdom.
1,999 units