FDA Enforcement Class II Terminated

Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.

Recall: Z-2435-2020 · Reported July 8, 2020

Enforcement

Recall Number
Z-2435-2020
Event ID
85576
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Westmed, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 8, 2020
Initiation Date
January 15, 2019
Classification Date
June 26, 2020
Termination Date
February 5, 2021
Address
5580 S Nogales Hwy Ste 170, N/A, Tucson, AZ, 85706-3333, United States

Description

Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.

Reason

A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Code Info

101618U02, 102318U06, 101518U03.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.

Quantity

35 cases