FDA Enforcement Class I Ongoing

WATCHMAN TruSeal Access System DBL, US, Material Number (UPN) M635TU70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Recall: Z-2432-2025 · Reported September 10, 2025

Enforcement

Recall Number
Z-2432-2025
Event ID
97363
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 10, 2025
Initiation Date
July 29, 2025
Classification Date
August 29, 2025
Address
1 Scimed Pl, N/A, Maple Grove, MN, 55311-1565, United States

Description

WATCHMAN TruSeal Access System DBL, US, Material Number (UPN) M635TU70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.

Reason

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

Code Info

GTIN 08714729965718, ALL NON-EXPIRED BATCHES

Distribution

Worldwide.

Quantity

340185 units in total