WATCHMAN TruSeal Access System ANT, OUS, Material Number (UPN) M635TS70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
Enforcement
- Recall Number
- Z-2428-2025
- Event ID
- 97363
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 10, 2025
- Initiation Date
- July 29, 2025
- Classification Date
- August 29, 2025
- Address
- 1 Scimed Pl, N/A, Maple Grove, MN, 55311-1565, United States
Description
WATCHMAN TruSeal Access System ANT, OUS, Material Number (UPN) M635TS70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
GTIN 08714729965756, ALL NON-EXPIRED BATCHES
Worldwide.
340185 units in total