FDA Enforcement
Class II
Terminated
12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA. USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS.
Recall: Z-2427-2012
·
Reported September 26, 2012
Enforcement
- Recall Number
- Z-2427-2012
- Event ID
- 63089
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smith & Nephew Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 26, 2012
- Initiation Date
- August 22, 2012
- Classification Date
- September 20, 2012
- Termination Date
- February 27, 2013
- Address
- 1450 E. Brooks Rd, Memphis, TN, 38116, United States
Description
12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA. USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS.
Reason
Modular necks were manufactured out of specification by a vendor; the taper provide is undersized.
Code Info
Lot Numbers: 12FTM0017A, 12FTM0017B, 12FTM0017C, 12FTM0019A. 12FTM0019B, 12FTM0019C, 12FTM0019D, 12FTM0020A, 12FTM0020B, 12FTM0020C, 12FTM0021A, 12FTM0021B, and 12FTM0021C
Distribution
Worldwide Distribution-USA (nationwide) including the states of CA, FL, GA, IL, IN, MA, NC, NJ, NY, OH, OK, PA, TN, TX, and WA, and the country of Canada.
Quantity
70 units