FDA Enforcement
Class II
Ongoing
Snowden-Pencer MIS DIAMOND-TOUCH CLAMP BOWEL 60 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7046; endoscopic surgical device
Recall: Z-2422-2023
·
Reported August 23, 2023
Enforcement
- Recall Number
- Z-2422-2023
- Event ID
- 92561
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Carefusion 2200 Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- August 23, 2023
- Initiation Date
- July 13, 2023
- Classification Date
- August 17, 2023
- Address
- 75 N Fairway Dr, Vernon Hills, IL, 60061-1845, United States
Description
Snowden-Pencer MIS DIAMOND-TOUCH CLAMP BOWEL 60 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7046; endoscopic surgical device
Reason
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Code Info
UDI/DI 10885403160578, Lot Numbers: C23, L22
Distribution
US and Canada
Quantity
4 units