Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2186 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: N/A
Enforcement
- Recall Number
- Z-2419-2025
- Event ID
- 97297
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- THOR Photomedicine Ltd
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 3, 2025
- Initiation Date
- July 24, 2025
- Classification Date
- August 22, 2025
- Address
- 2 Anglo Office Park, Units 1 &; 67 White Lion Road, Amersham, N/A, United Kingdom
Description
Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2186 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: N/A
The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
Model/Catalog Number: S2186; UDI-DI: Serial numbers: 4005, 4062;
Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.
2 units (1 US, 1 OUS)