FDA Enforcement Class II Terminated

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.

Recall: Z-2418-2018 · Reported July 18, 2018

Enforcement

Recall Number
Z-2418-2018
Event ID
80379
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Visit
Report Date
July 18, 2018
Initiation Date
June 21, 2018
Classification Date
July 10, 2018
Termination Date
June 16, 2021
Address
8200 Coral Sea St Ne, Mounds View, MN, 55112-4391, United States

Description

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.

Reason

Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.

Code Info

Product UPN/GTIN: 00885074300245 Serial Numbers: PNT402938H PNT413911H PNT413995H

Distribution

Worldwide Distribution - US Nationwide

Quantity

3 units