FDA Enforcement Class II Terminated

Radiometer Fixation Kit N20 for the TCM CombiM monitoring system, REF 905-873. The Fixation Kit N20 is used to apply tc Sensor 84 and tc Sensor 54 on the skin of a patient. TC sensor 84 and 54 are used with the TCM CombiM monitoring system and the Philips TcG10 module. The TCM CombiM monitoring system is intended for continuous transcutaneous monitoring of carbon dioxide (tcpCO2) and oxygen (tcpO2) partial pressures.

Recall: Z-2417-2015 · Reported August 26, 2015

Enforcement

Recall Number
Z-2417-2015
Event ID
71739
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Radiometer America Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 26, 2015
Initiation Date
July 16, 2015
Classification Date
August 19, 2015
Termination Date
January 29, 2016
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232, United States

Description

Radiometer Fixation Kit N20 for the TCM CombiM monitoring system, REF 905-873. The Fixation Kit N20 is used to apply tc Sensor 84 and tc Sensor 54 on the skin of a patient. TC sensor 84 and 54 are used with the TCM CombiM monitoring system and the Philips TcG10 module. The TCM CombiM monitoring system is intended for continuous transcutaneous monitoring of carbon dioxide (tcpCO2) and oxygen (tcpO2) partial pressures.

Reason

When the fixation ring is taken off the paper liner there is a potential for fractions of the paper liner's surface coating to stick to the surface of the ring.

Code Info

Model #: 905-873 Lots: 407221, 407231, 407241, 407251, 408041, 408191, 408201, 408211, 408231, 408232, 408261, 408271, 409021, 409101, 409231, 409291, 410021, 410141, 410281, 411171, 411201, 411261, 412021, 412111, 412161, 412201, 501081, 501131, 501191, 501311, 502031, 502101, 502131, 502171, 502231, 502261, 503091, 503231, 503311, 504031, 504081, 504111, 504271, 505041, 505061, 505141, 505181

Distribution

Worldwide Distribution - US Distribution and to the countries of : Canada, Spain, France, Great Britain, Philippines, Netherlands, Japan, China, Australia, Denmark, South Africa, Germany, Qatar, Turkey, Hong Kong, Costa Rica, Austria, Finland, Indonesia, Norway, Malaysia, Belgium, Czech Republic, Hungry, United Arab Emirates, Poland, Colombia, Belgium and Malta.

Quantity

4244 kits