FDA Enforcement Class II Ongoing

Brand Name: THOR Product Name: NovoTHOR Model/Catalog Number: S2188 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: No

Recall: Z-2416-2025 · Reported September 3, 2025

Enforcement

Recall Number
Z-2416-2025
Event ID
97297
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
THOR Photomedicine Ltd
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 3, 2025
Initiation Date
July 24, 2025
Classification Date
August 22, 2025
Address
2 Anglo Office Park, Units 1 &; 67 White Lion Road, Amersham, N/A, United Kingdom

Description

Brand Name: THOR Product Name: NovoTHOR Model/Catalog Number: S2188 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary increase in local blood circulation, temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, muscle spasm, and temporary relaxation of muscles. Component: No

Reason

The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.

Code Info

Model/Catalog Number: S2188; UDI-DI: 05060494130503; Serial Numbers: 4004, 4015, 4017, 4019, 4021, 4022, 4024, 4026, 4028, 4029, 4035, 4036, 4037, 4042, 4043, 4044, 4046, 4047, 4049, 4054, 4055, 4059, 4066, 4071, 4072, 4073, 4074, 4076, 4077, 4078, 4079, 4082, 4083, 4084, 4085, 4086, 4087, 4088, 4090, 4091, 4094, 4095, 4097, 4099, 4101, 4105, 4106, 4111, 4112, 4114, 4116, 4117, 4121, 4123, 4124, 4125, 4126, 4130, 4140, 4141, 4143, 4144, 4151, 4153, 4157, 4158, 4009, 4020, 4103, 4107, 4109, 4110, 4113, 4120, 4122, 4127, 4133, 4134, 4138, 4139;

Distribution

Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.

Quantity

80 units (66 US, 14 OUS)