FDA Enforcement Class I Terminated

Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA, Product Code IAP-0700

Recall: Z-2416-2020 · Reported July 8, 2020

Enforcement

Recall Number
Z-2416-2020
Event ID
85684
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Arrow International Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 8, 2020
Initiation Date
May 20, 2020
Classification Date
June 26, 2020
Termination Date
January 7, 2023
Address
2400 Bernville Rd, N/A, Reading, PA, 19605-9607, United States

Description

Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA, Product Code IAP-0700

Reason

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Code Info

Serial Number Ranges: Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA IAP-0700 150312F through 150313F 150315F 150318F 150321F 150324F through 150326F 161101F through 161110F 161112F through 161120F 161201F through 161210F 170101F through 170120F 170701F through 170710F 170801F through 170810F 170901F through 170920F 170931F through 170940F 171002F through 171010F 171031F through 171036F 171126F through 171130F 180101F through 180105F Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA IAP-0700 180121F through 180125F 180201F through 180210F 180311F through 180330F 180401F through 180410F 180421F through 180425F 180431F through 180440F 180511F through 180520F 180601F through 180605F 180701F through 180710F 180801F through 180810F 180821F through 180825F 180922F through 180931F 181041F through 181050F 181101F through 181110F 181201F through 181210F 190101F through 190110F 190131F through 190135F 190201F through 190210F 190231F through 190250F 190301F through 190303F 190305F through 190320F 190331F through 190335F 190401F through 190410F 190421F through 190430F 190441F through 190450F 190501F through 190510F 190541F through 190550F 190601F through 190620F 190641F through 190645F 190713F through 190730F 190801F through 190812F 190826F through 190834F 190901F through 190912F 190925F through 190930F 191013F through 191018F 191037F through 191042F 191049F through 191054F 191113F through 191136F 191201F through 191212F 200101F through 200106F 200137F through 200148F 200201F through 200212F Arrow AC3 Optimus Intra-Aortic Balloon Pump AC3 Optimus IABP NA/EMEA IAP-0700 200301F through 200306F 200325F through 200326F 200328F through 200339F 200341F through 200348F

Distribution

Worldwide distribution.

Quantity

2123 total