FDA Enforcement Class I Terminated

Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535

Recall: Z-2415-2020 · Reported July 8, 2020

Enforcement

Recall Number
Z-2415-2020
Event ID
85684
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Arrow International Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 8, 2020
Initiation Date
May 20, 2020
Classification Date
June 26, 2020
Termination Date
January 7, 2023
Address
2400 Bernville Rd, N/A, Reading, PA, 19605-9607, United States

Description

Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535

Reason

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Code Info

Serial Number Range: AEROAUTOCAT 2 WAVE IAP-0535 150907W 151030W 151101W 151104W 151106W 160305W through 160306W 160743W 160824W 190119W through 190120W 190221W 190223W through 190226W 191047W

Distribution

Worldwide distribution.

Quantity

2123 total