FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.
Enforcement
- Recall Number
- Z-2411-2018
- Event ID
- 80293
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Fort Defiance Industries, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- July 18, 2018
- Initiation Date
- May 24, 2018
- Classification Date
- July 10, 2018
- Termination Date
- June 27, 2023
- Address
- 2411 Maremont Pkwy, Loudon, TN, 37774-1066, United States
Description
FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641 Product Usage: The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.
During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.
Serial Numbers: AFS-0001 through AFS-0199
US Nationwide Distribution in the states of PA. VA. TX
199 units