FDA Enforcement Class III Terminated

ADVIA Centaur FSH Assay: ADVIA Centaur Calibrator B US: Calibrator B 00652707 (6 pack) or 00649625 (2 pack ) OUS:OUS: 2 pack REF: 00650577 and Ref: 00654130 (6 pack) Product Usage: ADVIA Centaur CP FSH assay: For in vitro diagnostic use in the quantitative determination of follicle-stimulating hormone (FSH) in serum using the ADVIA Centaur and ADVIA Centaur XP systems

Recall: Z-2410-2012 · Reported September 26, 2012

Enforcement

Recall Number
Z-2410-2012
Event ID
62897
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 26, 2012
Initiation Date
August 7, 2012
Classification Date
September 20, 2012
Termination Date
February 26, 2014
Address
333 Coney St, East Walpole, MA, 02032-1516, United States

Description

ADVIA Centaur FSH Assay: ADVIA Centaur Calibrator B US: Calibrator B 00652707 (6 pack) or 00649625 (2 pack ) OUS:OUS: 2 pack REF: 00650577 and Ref: 00654130 (6 pack) Product Usage: ADVIA Centaur CP FSH assay: For in vitro diagnostic use in the quantitative determination of follicle-stimulating hormone (FSH) in serum using the ADVIA Centaur and ADVIA Centaur XP systems

Reason

ADVIA Centaur Prolactin Assay and FSH, Calibrator B fails the analytical sensitivity stated in the IFU for lots below 40

Code Info

US: 2 Pack Ref: 00649625: Lot Numbers/Exp Date: Dec 2, 2012: 78813A30 79982A30 81013A30 82021B30 82767B30 83453C30 85447D30 87319A40 Exp Date:April 9, 2013 85598A40 85708A40 86098A40 87319A40 87979B40 88809B40 89351C40 91750C40 (6 pack) Ref. 00652707: Lot Numbers/Exp Date: Dec 2 2012 79354A30 80802A30 81892B30 82815B30 83741D30 84186A40 Exp. Date April, 8, 2013 85599A40 86191A40 88102B40 89717C40 91383C40 (6 Pack) Ref: 00654254 79714A30 Exp Date Dec 2, 2012 82370A30 84274A30 85512E30 OUS: 2 pack REF: 00650577 78319A29 Exp Dec 2, 2102 78487A29 79024A29 79403A29 79770B29 79981B29 82935B29 85735C29 00654130 Ref: 00654130 (6 pack) 78321A29 Exp Dec 2,2102 79197A29 83314C29

Distribution

Worldwide Distribution - US Nationwide and the countries of: Canada, Argentina, Australia, Bangaldesh,Brazil, Chile, China, Columbia, Croatia, Ecuduar, Egypt, Hong Kong, India, Israel,Japan, Korea, Maylasia, New Zealand, Mexico, Peru, Philipines, Pakistan, Saudia Arabia, Singapore, Thailand, Taiwan, Viet Nam, South Africa and UK.

Quantity

15,445 kits