FDA Enforcement Class II Ongoing

Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Recall: Z-2406-2021 · Reported September 8, 2021

Enforcement

Recall Number
Z-2406-2021
Event ID
88451
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Elekta Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 8, 2021
Initiation Date
July 26, 2021
Classification Date
September 2, 2021
Address
1450 Beale St Ste 205, N/A, Saint Charles, MO, 63303, United States

Description

Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Reason

If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.

Code Info

Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10 UDI: (01)00858164002190(10) 5.40.00 (01)00858164002190(10) 5.40.01 (01)00858164002190(10) 5.40.02 (01)00858164002268(10) 5.51.10

Distribution

Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom

Quantity

N/A