FDA Enforcement Class II Ongoing

Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

Recall: Z-2405-2026 · Reported June 17, 2026

Enforcement

Recall Number
Z-2405-2026
Event ID
98963
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abbott Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 17, 2026
Initiation Date
May 12, 2026
Classification Date
June 11, 2026
Address
15900 Valley View Ct, N/A, Sylmar, CA, 91342-3577, United States

Description

Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

Reason

Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.

Code Info

Programmer Model 3650; Software Model 3330; GTIN 05414734509725; Software Versions 25.0.1 - 28.7.1 in the United States (varies based on region)

Distribution

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet

Quantity

11182 programs