FDA Enforcement Class II Terminated

Precise Digital Accelerator Delivery of radiation to defined target volumes

Recall: Z-2405-2012 · Reported September 26, 2012

Enforcement

Recall Number
Z-2405-2012
Event ID
62917
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Elekta, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
September 26, 2012
Initiation Date
August 6, 2012
Classification Date
September 19, 2012
Termination Date
October 8, 2014
Address
4775 Peachtree Industrial Blvd, Bldg 300, #300, Norcross, GA, 30092-3011, United States

Description

Precise Digital Accelerator Delivery of radiation to defined target volumes

Reason

There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.

Code Info

151587, 105932, 151574, 151363, 151076, 151110, 151369, 151504, 105941, 151418, 151674, 151566, 151259, 151636, 151568, 151390, 151258, 151584, 151391, 151317, 151522, 151055, 151423, 151507, 151623, 151627, 151640, 151479, 151340, 151178, 151467, 151514, 151051, 151411, 151216, 151430, 151173, 151616, 151595, 151141, 151433, 151611, 151151, 151653, 151515, 151360, 151675, 151538, 151024, 151523, 151521, 151260, 151519, 151432, 151410, 151371, 151084, 151435, 151387, 151530, 151328, 151176, 151080, 151626, 151567, 151598, 151601, 151269, 151396, 151557, 151558, 105982, 151160, 151634, 151394, 151398, 151540, 151632, 151446, 151301, 151038, 151441, 151256, 151168, 151156, 151157, 151155, 151579, 151412, 105984, 151053, 151359, 151585, 151326, 151327, 151443, 151250, 151130, 151245, 151416, 151664, 151672, 151074, 151352, 151298, 151545, 151628, 105803, 151362, 151073

Distribution

Nationwide Distribution including AL, AZ, CA, CO, CT, FL, GA, IL, IA, MD, MA, MI, MN, NE, NV, NJ, NM, NY, ND, OH, OK, OR, PA, VA, SC, TN TX, WA, WV, WI and Puerto

Quantity

110