OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.
Recall: Z-2403-2016
·
Reported August 24, 2016
Enforcement
- Recall Number
- Z-2403-2016
- Event ID
- 74843
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- OraSure Technologies, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 24, 2016
- Initiation Date
- July 20, 2016
- Classification Date
- August 17, 2016
- Termination Date
- May 10, 2017
- Address
- 1745 Eaton Ave, N/A, Bethlehem, PA, 18018-1769, United States
Description
OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.
Reason
Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.
Code Info
Product Code 3001-2807 and the lot number is 6648965. Expiration date October 31, 2016.
Distribution
Distributed to one customer only in US state of Georgia.
Quantity
3,075 units