FDA Enforcement Class I Terminated

OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.

Recall: Z-2403-2016 · Reported August 24, 2016

Enforcement

Recall Number
Z-2403-2016
Event ID
74843
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
OraSure Technologies, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 24, 2016
Initiation Date
July 20, 2016
Classification Date
August 17, 2016
Termination Date
May 10, 2017
Address
1745 Eaton Ave, N/A, Bethlehem, PA, 18018-1769, United States

Description

OraQuick Ebola Rapid Antigen Test, Product Code 3001-2807; Intended for the qualitative detection of antigens for Ebola viruses.

Reason

Failed stability testing at 8 months. This lot produced intermittent false negative results for the positive kit control with devices stored at 30C conditions.

Code Info

Product Code 3001-2807 and the lot number is 6648965. Expiration date October 31, 2016.

Distribution

Distributed to one customer only in US state of Georgia.

Quantity

3,075 units