FDA Enforcement
Class II
Terminated
Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a rest surface with adjustable height from floor and adjustable surface contours.
Recall: Z-2402-2016
·
Reported August 17, 2016
Enforcement
- Recall Number
- Z-2402-2016
- Event ID
- 73577
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Invacare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 17, 2016
- Initiation Date
- March 15, 2016
- Classification Date
- August 11, 2016
- Termination Date
- September 15, 2017
- Address
- 1200 Taylor St, Elyria, OH, 44035-6248, United States
Description
Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a rest surface with adjustable height from floor and adjustable surface contours.
Reason
The firm recently identified a quality issue with the Junction box used in the Invacare Bariatric Bed. They discovered that the Junction Box may omit sparks which could result in a fire in the presence of flammable materials.
Code Info
Model #'s:1104343 & 1104345
Distribution
US Nationwide Distribution including DC and PR. Internationally Canada.
Quantity
1,011 units